Spirometry: update of the procedure and post pandemic perspectives

Benítez-Pérez, Rosaura Esperanza1; Cortés-Telles, Arturo2; Meneses-Tamayo, Érika1; Silva-Cerón, Mónica1; del Río-Hidalgo, Rodrigo Francisco3,4; Monraz-Pérez, Sergio1; González-Castellanos, Mabelle1; Camargo-Ángeles, Robert4; Torre-Bouscoulet, Luis5; Vázquez-García, Juan Carlos1; Pérez-Padilla, Rogelio1; Gochicoa-Rangel, Laura1,5
Full text How to cite this article 10.35366/115397

DOI

DOI: 10.35366/115397
URL: https://dx.doi.org/10.35366/115397

Language: English/Spanish [Versi?n en espa?ol]
References: 44
Page: 104-124
PDF size: 1976.08 Kb.


Key words:

spirometry, lung function, vital capacity, bronchial obstruction.

ABSTRACT

Spirometry is the most widely used and standardized respiratory function test. In 2019, the American Thoracic Society and the European Respiratory Society updated the international guidelines for its execution. The COVID-19 pandemic forced the establishment of better biosafety parameters and has renewed interest in respiratory medicine in the world, including physiological evaluation. The present manuscript summarizes these changes incorporating recommendations and suggestions for countries with limited resources.



ABBREVIATIONS:

  • ATS/ERS = American Thoracic Society/European Respiratory Society.
  • COVID-19 = disease by coronavirus 2019.
  • EOFE = end of forced expiration.
  • EOTV = end of test volume.
  • F/V = flow/volume.
  • FET = forced expiration time.
  • FEV1 = forced expiratory volume in one second.
  • FVC = forced vital capacity.
  • FIVC = forced inspiratory vital capacity.
  • IC = inspiratory capacity.
  • PEF = peak expiratory flow.
  • RV = residual volume.
  • SVC = slow vital capacity.
  • TLC = total lung capacity.
  • V/T = volume/time.



INTRODUCTION

Forced spirometry is a respiratory function test that assesses components of lung mechanics. It measures the maximum volume of air that an individual can forcibly inhale and exhale as a function of time.1-3

The main variables measured in forced spirometry are: forced vital capacity (FVC), which is the maximum volume of air, measured in litres, that can be exhaled through the mouth with maximum effort after a full inspiration; and forced expiratory volume in one second (FEV1), which is the volume of air exhaled during the first second of the FVC manoeuvre.1-3

The FEV1/FVC ratio, expressed as a percentage of the absolute value, is the defining variable for obstruction, as it represents a disproportionate reduction in the maximum FEV1 of a forced manoeuvre relative to the total volume that a subject can exhale during the FVC manoeuvre. Therefore, a decreased FEV1/FVC ratio implies airflow limitation, i.e. airway obstruction during exhalation.2

Other spirometric variables briefly addressed in this paper are FEV0.5 and FEV0.75, which represent the forced expiratory volume in the specified fraction of a second,1,4 as well as the slow vital capacity (SVC) which is the maximum volume of air measured in litres, that can be exhaled through the mouth in a relaxed manner after a full inspiration.1

This document aims to adjust the spirometry procedure according to the new recommendations published in the 2019 ATS/ERS standard for forced spirometry,1 as well as the challenges and protocols assumed in its execution in the wake of the COVID-19 pandemic.5



INDICATIONS

Spirometry is a fundamental test in respiratory assessment.1 Indications are shown in Table 1, updated according to current evidence on the usefulness of the test in multiple settings.

CONTRAINDICATIONS1

The FVC manoeuvre directly increases intrathoracic pressure and indirectly increases intra-abdominal and intracranial pressure, and physical exertion can increase myocardial demand. Therefore, the potential risks when performing forced spirometry are due to the impact of these changes on thoracic organs, abdominal, venous return and arterial pressure. The international standard for spirometry1 states that there are no absolute contraindications for performing forced spirometry and that its performance will depend on the risk-benefit of performing the test.

Patients with contraindications should ideally be evaluated in a pulmonary function laboratory by expert personnel. Slow spirometry may sometimes be performed to assess the subject's vital capacity, although the result is not interchangeable. It is also important that the patient is cooperative and follows instructions in order to avoid submaximal manoeuvres. The appearance of pain, presyncope or discomfort during the test is a criterion for suspension of the test.

Contraindications, all relative, are shown in Table 2.



MATERIAL RESOURCES



SPIROMETER

A device that records physiological ventilatory volumes, in the vital capacity range, as well as the flow generated by them through a sensor.27

ATS/ERS 20191 states that this equipment must meet the minimum requirements of the latest update to ISO: 26782,27 which are summarised below:

  • 1. Measuring range from 0 to 8 litres, under BTPS (Body Temperature and Pressure and Saturated) conditions.
  • 2. Maximum permissible error in volume measurement of ± 3% or 0.050 L (whichever is greater).
  • 3. Minimum expiratory time recording of 15 seconds.
  • 4. Real-time graphs with a ratio for the F/V (flow/volume) graph of 2 L/s: 1 L and for the V/T (volume/time) graph of 1 L: 1 s.
  • 5. Recording of the extrapolated volume as well as the volume at the end of the forced exhalation (to identify start and end criteria).
  • 6. Impedance of the equipment, with all its accessories, less than 0.15 kPa/(L/s) with flow rates up to 14 L/s.
  • 7. Have a weather station for temperature measurement, which must have an accuracy of ± 1 °C to properly calculate the BTPS correction factor; in case of not having a weather station, the calculation of the correction factor to BTPS units will have to be done manually.
  • 8. Appropriate reference equations for the population.3
  • 9. Real-time display of F/V and V/T graphs at the time of manoeuvres.1
  • 10. The report generated should have both F/V and V/T graphs for each of the manoeuvres performed; a volume scale ≥ 10 mm/L and a time scale ≥ 20 mm/s is recommended.3

Volumetric spirometers have fallen into disuse and flow spirometers, which measure air displacement velocity and calculate volume by integration, now predominate. These spirometers are portable, easy to clean, and some use disposable sensors, reducing the risk of cross-contamination; features that have facilitated the incorporation of spirometry in the office, hospital, laboratory and even the patient's home.

Table 3 summarises the different types of flow spirometers, their advantages and disadvantages, virtually all of which are available in Mexico.

Other equipment and consumables3

  • 1. Computer and printer (some equipments do not require).
  • 2. Scale, stadiometer.
  • 3. Stable chair with side armrests. Avoid chairs with wheels to prevent falls.
  • 4. Room thermometers with an accuracy of 1 °C and hygrometer for relative humidity measurement.
  • 5. Mouthpieces recommended by the manufacturer, diving mouthpieces can be used for those patients who are unable to make a good lip seal.
  • 6. Nasal forceps.
  • 7. Certified three-litre syringe.



INFECTION CONTROL SUPPLIES:

  • 1. Access to hand washing and gel-alcohol.
  • 2. Disposable in-line filters with > 99% efficiency for filtration of viruses, bacteria and mycobacteria; dead space < 100 mL and resistance less than 1.5 cm H2O at 6 L/s flow rate.
  • 3. N95 respirator with leakage of less than 10% and a filtration efficiency of > 95% at a flow rate of 50 L/min.
  • 4. Protective eyewear.
  • 5. Natural water should be available, as well as facial tissues to be offered to the patient in case of coughing or secretions.



BRONCHODILATOR CONSUMABLES:

  • 1. Salbutamol in metered dose inhaler (100 mg by atomisation).
  • 2. Ipratropium bromide aerosol (20 mg per atomisation).
  • 3. Reservoir chamber (spacer) with a recommended volume of at least 300 mL.

QUALITY CONTROL AT THE WORKPLACE (Figure 1)

There are requirements that any laboratory or site, where spirometry is performed, must meet to ensure good practice.

Logbook2

We recommend an auditable quality control report. The log should include the results of calibration processes, dates of spirometer maintenance (repairs and/or adjustments) and calibration syringe, software/hardware updates (dates on which procedures were performed and check that reference equations are included). Logbooks should be under the custody of the responsible technical professional and always available. Each spirometer should have its own physical or electronic logbook.



EVALUATION OF SPIROMETER ACCURACY

All spirometry equipment must be validated prior to release and subsequently calibrated or verified (before manoeuvres are performed) to meet accurate flow and volume measurements for consistent clinical decisions. The following is a breakdown of the concepts describing these processes:

Validation: assesses the reproducibility of expiratory manoeuvres using a computerized curve generation system. The spirometer must faithfully reproduce the model curve generated by the computer system. It is not part of standard laboratory procedures; it is performed before spirometers are marketed.28

Calibration: a procedure by which a relationship is established between the volume or flow measured by the sensor and the actual flow or volume of the calibrator (syringe) under Ambient Temperature, atmospheric Pressure, Saturated (ATPS) conditions. It is an electrical gain adjustment manoeuvre of the device.1 Spirometers whose sensor is affected by gas characteristics (condensation, etc.) such as pneumotachographs should always be calibrated.

Calibration verification: procedure that verifies that the spirometer is within the accuracy limits (± 3%, corresponding to: ± 2.5% of the equipment and ± 0.5% of the calibration syringe) under ATPS conditions. If the device fails, the verification must be repeated and the possible causes (leakage at the syringe junction with the spirometer, zero flow error, error in the syringe filling and injection process, or malfunction of the syringe) must be analyzed. If the failure is repeated, the equipment should be sent for recalibration and maintenance. The procedure should be performed daily by discharging the calibration instrument (certified three-litre syringe) through the sensor at least three times, in a flow range between 0.5 to 12 L/s (with injection times between 0.5 to 6 seconds). The final result should yield a volume of 3 L ± 90 mL (± 3%). If spirometry is intended to be performed with filters, then this procedure should also use filters. There are spirometers that are pre-calibrated by the manufacturer and cannot be recalibrated by the spirometer operator, however, it is advisable to carry out the calibration verification process on all equipment (even ultrasonic) and generate the corresponding reports that will be filed in the logbook.1

Biological control: performed with the participation of a healthy pulmonary subject, without risk factors, with normal spirometry; usually a laboratory technician who has the ability to perform the procedure in a highly repeatable manner. It does not replace the use of the calibration syringe. The subject to be designated as the biological control is given serial spirometry for a specified number of days at the same time, averaging the highest FEV1 and FVC values.1 Some machines provide the option of automated calculation of the SDs (standard deviations) within which the measurements obtained are considered to be correct. Subsequently, the subject should perform weekly spirometry to compare the measured value against the maximum expected error.

Standardized calibration instrument1

Corresponds to a three-litre syringe with an accuracy of ± 15 mL or ± 0.5%, with a current certificate. It should be checked according to the manufacturer's recommendations and ideally once a month, looking for possible leaks, performing a manoeuvre that attempts to fill and empty the syringe (at different volumes) with the outlet blocked. Similarly, it should be stored away from moisture or heat. A damaged or knocked syringe is considered potentially out of calibration and should not be used for this procedure.

Procedure conditions1-3,28

  • 1. ATPS: ambient temperature, barometric pressure and ambient water vapour saturation. All procedures involving the input and output of volumes from a calibration syringe to the sensor must be performed under these conditions.
  • 2. BTPS correction: given that exhaled air is at a body temperature of 37 °C and saturated with a water vapour pressure of 47 mmHg; the partial pressure in the lungs is 760-47 mmHg = 713 mmHg (at sea level). All procedures involving the performance of manoeuvres on subjects must conform to these conditions.
  • 3. The work site should ideally meet the following characteristics: temperature between 17 to 35 °C, relative humidity between 30 to 75%. Changes in temperature or humidity during the working day should be recorded in the logbook as this may be a source of variability in spirometry results. In situations where the ambient air temperature changes rapidly > 3 °C in < 30 min, the appropriate correction should be made in the spirometer. Artificially air-conditioned laboratories allow better control of environmental variables.
  • 4. Some spirometers have built-in sensors that automatically measure temperature and barometric pressure, but it is recommended that the operator verify the accuracy of these parameters.1



ONGOING STAFF TRAINING

Staff performing spirometry should maintain competence through regular training to safeguard the quality of results. Lack of ongoing training and infrastructure contribute to lack of knowledge about this test.29 A short course improves competence;30 learning is reinforced by a second training of longer duration and close monitoring.29 It is recommended that staff develop skills to cope with special situations including: non-English language (dialects), hearing or visual impairments, and uncooperative patients.1

Quality of manoeuvres. Review the relevant section.

Improving patient experience31

The European Lung Foundation (ELF) conducted a virtual survey in 2018 in 52 countries among patients who regularly underwent spirometry. Of the 1,760 respondents, only 17% of them rated the test as difficult to perform, the rest rated it as tolerable. The most important suggestions from patients were the following: clear and concise information before, during and after the test (regarding drug cessation, contraindications, etc.), as well as access to and explanation of the results obtained in the context of their pathology.

FORCED SPIROMETRY PROCEDURE WITH BRONCHODILATOR1,3



RECOMMENDATIONS FOR THE PATIENT PRIOR TO THE TEST

  • 1. No smoking, vaping, or using water pipes at least one hour beforehand.
  • 2. No use of drugs that affect consciousness within eight hours before.
  • 3. No strenuous exercise one hour before.
  • 4. Avoid wearing restrictive chest or abdomen garments.
  • 5. If the indication for the test is diagnostic, bronchodilators should be discontinued according to the time of action of each (Table 4).



PREPARATION OF EQUIPMENT PRIOR TO TESTING

  • 1. All components (hoses, sensors, connectors, etc.) must be properly disinfected and/or sterilised and assembled according to the manufacturer's instructions.
  • 2. Perform calibration or calibration verification.
  • 3. The spirometer must be coded to the altitude or barometric pressure and average relative humidity of the site where the study is performed.
  • 4. Verify that the spirometry report is properly configured.



STAFF ACTIONS ON ARRIVAL OF THE PATIENT

  • 1. Introduce yourself to the patient and check that your details are correct (check name and date of birth).
  • 2. Review the indications.
  • 3. Evaluate the presence of possible contraindications, vital signs and the patient's adherence to the recommendations.
  • 4. Enter patient data into the spirometer: full name, date of birth, anthropometric parameters: age in years at the day of testing, sex at birth (patients may provide their gender identity, but should be informed that sex at birth is required as it is a determinant of predicted lung size). It is important to enter, as notes or observations, any additional data that may help in the further interpretation of the study (such as smoking, exposures, history of previous lung disease, etc.).
  • 5. Obtain the weight in light clothing, on a precision scale and record it in kilograms in closed units to the nearest 0.5 kg.
  • 6. Obtain height with a stadiometer (should be measured without shoes, with feet together, standing as upright as possible, facing forward, with back and heels against the wall or stadiometer.)
  • 7. In patients unable to stand or with rib cage deformity, measurement of arm span may be used to estimate standing height; measure the distance between the tips of the middle fingers (wingspan). For Caucasian males: height = arm span/1.03, for African-American males: height = arm span/1.06 and for females height = arm span/1.01. For patients who cannot be measured standing and also do not have an arm, the mid-span can be measured as the distance between the tip of the middle finger and the prominent cervical vertebra. And in patients with significant body posture deformity in whom it is not possible to measure the wingspan linearly, the composite wingspan shall be calculated.
  • 8. Place the patient in a seated position, in a chair without wheels and with arm support, with the chest and neck in an upright position. If the test is performed in a different position (e.g. decubitus) it should be recorded.
  • 9. Explain to the patient, in simple words, the purpose of the test. The following sentence is recommended: "Spirometry is a blowing test to measure the size of the lungs and to find out whether or not there is obstruction of the bronchial tubes. You are going to blow hard and steady through this mouthpiece several times until you get at least three proper manoeuvres".
  • 10. A video or picture can be used to reinforce the explanation. It is not advisable to remove eye or respiratory protection to demonstrate the manoeuvre.



FORCED VITAL CAPACITY (FVC) MANOEUVRES

Manoeuvres that assess both phases of the respiratory cycle (inspiration and expiration), also known as "closed circuit", are preferred.

The four basic steps of a good manoeuvre are as follows:

  • 1. Maximal inspiration to total lung capacity (TLC).
  • 2. Explosive, immediate and unhesitant exhalation.
  • 3. Continuing exhalation until the end of forced expiration criteria are met. At this point the motivation given by the performer is important.
  • 4. Breathe in again until TLC. This closes the inspiratory curve and allows assessment of the forced inspiratory vital capacity (FIVC).

In spirometers that do not record the inspiratory phase, open circuit manoeuvres can be performed by placing the mouthpiece immediately after inspiration (step 1) and removing it after meeting end of forced expiratory criteria (step 3).

Figure 2 outlines the steps of a closed-circuit manoeuvre.

Bronchodilator administration32

The administration protocol should be recorded in writing in the site's internal procedures manual; it should contain the following elements:

  • 1. Type of bronchodilator (salbutamol, fenoterol, ipratropium bromide or combinations).
  • 2. Dose to be administered in children (salbutamol: 200 µg) and adults (salbutamol: 400 µg, ipratropium bromide: 80 µg).
  • 3. Method of administration: metered dose inhaler (MDI) with or without mask, nebuliser. In this regard, MDI with air chamber is recommended; the use of nebulisers should take into account air flow, equipment pressure and not use oxygen while administering the drug.
  • 4. Waiting time for post-bronchodilator manoeuvres: when using salbutamol wait at least 15 minutes and in the case of ipratropium bromide at least 30 minutes.

QUALITY OF MANOEUVRES1



EVALUATION OF EACH MANOEUVRE

At the end of each manoeuvre, it should be assessed for compliance with the following technical acceptability criteria:

  • 1. Adequate onset of forced expiration (Figure 3A). An explosive onset, with maximal effort, ensures that we are obtaining the patient's true FEV1.
  • The following two indicators must be assessed and met to ensure that the manoeuvre had a correct onset:

    • a. The extrapolated volume, which is the amount of gas that is exhaled in a hesitant manner from peak inspiration at time 0, should be < 5% of the FVC or.
    • b. The flow/volume (F/V) curve should be triangular in morphology, with steep, vertical rise to peak flow and gradual decline to 0.

  • 2. Adequate end of forced expiration (Figure 3B). A proper end of the manoeuvre ensures that the patient's true FVC is obtained.
  • At least one of the following three indicators must be assessed and met to ensure that the manoeuvre was properly terminated:

    • a. Plateau. This is the best indicator of the end of forced expiration; it refers to an absent increase in volume on the V/T or EOFE curve.
    • b. Expiratory time of 15 seconds. This indicator is more likely to be achieved in older adults or in patients with lower airway obstruction. For patient safety, if the plateau has not been reached, but 15 seconds of expiratory time has been reached, the manoeuvre should be terminated.
    • c. The subject is unable to continue exhaling. In this case, the FVC of the previous and subsequent manoeuvres shall be assessed to ensure that they are repeatable with respect to each other.

  • 3. Absence of artefacts affecting technical acceptability (Figures 4, 5, 6, 7, 8, 9, 10 and 11).

    • a. Baseline errors. Occurs when the operator or patient generates some flow while the equipment is establishing the baseline; affects both FEV1 and FVC (Figure 4). This artefact is common when spirometry is performed outdoors or near an air conditioning device.
    • b. High extrapolated volume. This is generated when the patient takes too long between maximal inspiration and expiratory effort. Its presence renders both FEV1 and FVC unacceptable and not useful (Figure 5).
    • c. Mouthpiece leak or obstruction. Occurs when the patient does not seal the mouthpiece properly with the lips, sticks the tongue in or bites down hard on the mouthpiece. If after verifying proper mouthpiece placement, subsequent manoeuvres still show the "artefact" of obstruction, it is important to rule out true intra- or extrathoracic central airway obstruction. Patients with facial paralysis or edentulous patients without prostheses may require support in sealing (Figures 6 and 7).
    • d. Coughing. If it occurs during the first second it affects the FEV1 result; however, FVC may be usable in such cases (Figure 8).
    • e. Glottic closure. The individual pushes instead of exhaling, closing the glottis and suddenly obstructing outflow; if it occurs during the first second it affects both FEV1 and FVC making the manoeuvre not usable, if it occurs after the first second FEV1 may be usable (Figure 9).
    • f. Repeated exhalations. Occurs when the patient reinhales through the nose and exhales again, falsely increasing the FVC of that manoeuvre. The FEV1 may be usable if the second exhalation occurred after the first second (Figure 10).
    • g. Variable efforts. Occurs when the patient does not exhale at maximal effort, which may be suspected in the presence of variable peak flows and non-overlapping traces on the F/V curve (Figure 11).
    • h. FIVC-FVC subtraction greater than 0.10 L (100 mL) or 5% of FVC (whichever is higher). If the volume inspired to close the circuit after completion of the forced expiration (FVIC) is much greater than the FVC for that manoeuvre, it means that the patient did not fully inspire at the start of the test, which affects the FVC result.

ASSESSMENT BETWEEN MANOEUVRES1

Acceptability implies that each manoeuvre is well executed (from start to finish), while repeatability means that the manoeuvres resemble each other, which is an indicator of measurement consistency. The more consistent a phenomenon is, the lower the probability of error.

Therefore, once three acceptable manoeuvres (both FEV1 and FVC) have been obtained, repeatability should be assessed under the following criteria: the difference between the two highest FEV1 values and the two highest FVC values should be ≤ 150 mL (maximum ≤ 200 mL) in subjects older than six years, and ≤ 100 mL or ≤ 10% of the highest value in patients younger than six years (Figure 12 and Table 5).

Test quality grades1

After obtaining three acceptable and two repeatable manoeuvres (in FEV1 and FVC) the quality of the test should be graded (Table 6). It is important to note that automated quality does not always match that of the expert observer, so automated algorithms should be used with caution.

Test quality considerations1

  • 1. Acceptability may be achieved in FEV1 but not in FVC, and vice versa, in the presence of certain artefacts that are difficult to correct for.
  • 2. Repeatability is analysed until three acceptable efforts have been completed.
  • 3. Eight attempts is a practical limit, but some people, especially those with little testing experience, may get their best manoeuvre after the eighth, especially if they do not show fatigue, in order to get three acceptable ones. Patients who are getting poorer tests or lower measurements with new manoeuvres will generally not benefit from going beyond the eighth manoeuvre.
  • 4. In case no acceptable manoeuvres in FEV1 and FVC are obtained, despite our and the patient's efforts, an expert assessor can use the quality score U and issue some interpretation.
  • 5. In patients with bronchial hyperresponsiveness, repeated FVC manoeuvres may cause decreased flows.

Reporting the results1-5

It is recommended that it includes sufficient information to assess the quality of the test, as well as a standardised interpretation by an expert. It should include the following components (Figure 12 and Table 5):

  • 1. Patient's full name.
  • 2. Patient's date of birth.
  • 3. Anthropometric parameters (age, gender, ethnicity, weight and height).
  • 4. Significant respiratory history.
  • 5. Origin of reference values.
  • 6. Date of last calibration.
  • 7. The values of three acceptable spirometry manoeuvres: FVC, FEV1 and PEF in units (L or L/s) to two decimal places and the FEV1/FVC ratio in percent to one decimal place. Depending on the case, FEV6 and FEV1/FEV6 can also be included, and in the case of preschoolers, FEV0.5 and FEV0.75, with absolute values and the percentage of the predicted value. It is essential to include the numerical value of the extrapolated volume (BEV, BEV or Vext) and end-expiratory volume (EOTV or EOFE), as well as the expiratory time (FET).
  • 8. All three volume-time and flow-volume graphs, both baseline and post-bronchodilator, should be visible.
  • 9. The three best values at baseline and the three best values after bronchodilator administration should be displayed. Some outdated spirometers report only the best baseline and best post-bronchodilator values.
  • 10. Ideally the change in FEV1 and FVC between the best baseline and best post-bronchodilator test should be reported.
  • 11. Column of predicted, lower limit of normal and Z-score.



BIOSAFETY CONSIDERATIONS

Precautions and processes that reduce the risk of personnel exposed to a potentially infectious agent33 have always been indispensable, but often ignored in healthcare procedures. Potential micro-organisms involved in cross-infection within a pulmonary function laboratory are mainly transmitted by droplets and aerosols; there is also a risk of contact transmission in immunocompromised patients.34

With the emergence of COVID-19, numerous expert consensus proposed limiting the performance of respiratory function tests according to pandemic phase.35-37 In a survey of laboratories registered with the American Thoracic Society, only 70% of them continued to perform spirometry in all phases.38 Although there is no similar information from Latin America, the National Institute of Respiratory Diseases in Mexico established biosafety guidelines that ensured the continuity of respiratory physiology services throughout the pandemic.39

Spirometry, like other respiratory function tests, is an aerosol-generating procedure, which requires active breathing manoeuvres in close proximity to the personnel performing the test and may induce coughing. In addition, asymptomatic and pre-symptomatic patients are difficult to detect, regardless of the screening performed.40 The pandemic itself has shown us how important it is to assess patients in the respiratory setting, and this includes spirometry.5,38 For full and urgent reactivation of centres performing this test, a hierarchy of risk control must be established, giving priority to processes that protect the collective workforce, without neglecting the surveillance of individual measures.41

In post-pandemic times, all hygiene and disinfection procedures should be documented in the internal procedures manual. Similarly, the scheduling of each study should consider extending the time between patients, thus avoiding prolonged waiting time and overcrowding of subjects, and reducing the risk of infection. Table 7 summarises the measures according to the international consensus published by the European Respiratory Society.



CONSIDERATIONS IN PAEDIATRICS

With proper training, children as young as two and a half years can perform acceptable spirometry.1

It is suggested that tidal volume manoeuvres (such as impulse oscillometry) be performed first, followed by forced spirometry, as deep inhalations may change bronchial tone in children with asthma. Children have a high elastic recoil, so these patients may not reach a plateau, in these cases the indicator of adequate EOFE is the repeatability of FVC. Practitioners involved in the performance of pulmonary function testing of young children should be trained to work with this population. Table 8 summarises some recommendations in this type of patient.42,43



SLOW SPIROMETRY

Spirometry can also be performed in a relaxed or quiet manner, which is referred to as slow spirometry. The slow manoeuvre is comfortable to perform, does not require strenuous physical effort and provides additional information to the forced manoeuvre.

The main measurements obtained from slow spirometry are SVC, which is the slowly exhaled volume from TLC to residual volume (RV), and inspiratory capacity (IC), which is the slowly inspired volume of air from expiration at tidal volume to maximal inspiration at TLC. It is recommended to be performed before any forced manoeuvre. The manoeuvre consists of the following steps (Table 9):

  • 1. Tidal volume breaths (at least three steady ones) and then ask the patient to perform one of the following:

  • a) Deep breath in to TLC without hesitation and relaxed exhalation to RV (Figure 13A).
  • b) Relaxed exhalation to RV and subsequent deep inspiration to TLC (Figure 13B).

The manoeuvres are relaxed and unforced. Peak inspiratory and expiratory levels are usually achieved within the first six seconds, some patients may require more time. The manoeuvre should not be excessively slow, as this may underestimate the SVC. As with forced spirometry, at least three acceptable manoeuvres should be obtained. A maximum of eight attempts is recommended. The criteria for acceptability are the same as for the forced manoeuvre. Repeatability is assessed by subtracting the values of the two best SVCs, and these should be < 150 mls.

The lack of repeatability in this study is, in most cases, due to incomplete inspiration. For SVC, the highest value of the acceptable manoeuvres should be reported. For IC, the average of the acceptable manoeuvres should be reported. SVC and CI are useful for assessing bronchodilator response, using as criteria for significant response an improvement in 200 mL and 12% in either variable.

FVC and SVC in subjects without airflow obstruction show similar values; however, in obstructed subjects, small airway collapse and air trapping in the forced manoeuvre mean that FVC is lower than SVC.

Limitations of slow spirometry are its lower reproducibility with respect to the forced manoeuvre, less standardisation and fewer reference values available; but, on the other hand, it may be more sensitive in detecting airflow obstruction.


REFERENCES

  1. Graham BL, Steenbruggen I, Miller MR, Barjaktarevic IZ, Cooper BG, Hall GL, et al. Standardization of spirometry 2019 update. An official American thoracic society and European respiratory society technical statement. Am J Respir Crit Care Med [Internet]. 2019;200(8):e70-e88. Available in: http://dx.doi.org/10.1164/rccm.201908-1590st

  2. Benítez PRE, Silva CM, Gochicoa RLG. Manual de procedimiento de espirometría forzada, departamento de fisiología respiratoria. Instituto Nacional de Enfermedades Respiratorias Ismael Cosío Villegas. México: INER, SSA; 2021.

  3. Benítez-Pérez RE, Torre-Bouscoulet L, Villca-Alá N, Del Río-Hidalgo RF, Pérez-Padilla R, Vázquez-García JC, et al. Espirometría: recomendaciones y procedimiento. Neumol Cir Torax [Internet]. 2016;75(2):173-190. Disponible en: http://dx.doi.org/10.35366/67124

  4. Beydon N, Davis SD, Lombardi E, Allen JL, Arets HGM, Aurora P, et al. An official American Thoracic Society/European Respiratory Society statement: pulmonary function testing in preschool children. Am J Respir Crit Care Med [Internet]. 2007;175(12):1304-1345. Available in: http://dx.doi.org/10.1164/rccm.200605-642st

  5. McGowan A, Laveneziana P, Bayat S, Beydon N, Boros PW, Burgos F, et al. International consensus on lung function testing during the COVID-19 pandemic and beyond. ERJ Open Res [Internet]. 2022;8(1):00602-2021. Available in: http://dx.doi.org/10.1183/23120541.00602-2021

  6. Vázquez-García JC, Hernández-Zenteno RJ, Pérez-Padilla JR, Cano-Salas MC, Fernández-Vega M, Salas-Hernández J, et al. Guía de Práctica Clínica Mexicana para el diagnóstico y tratamiento de la enfermedad pulmonar obstructiva crónica. GUÍA MEXICANA DE EPOC, 2020. Neumol Cir Torax [Internet]. 2019;78(Suppl: 1):4-76. Disponible en: http://dx.doi.org/10.35366/nts191a

  7. 2023 GOLD Report [Internet]. Global Initiative for Chronic Obstructive Lung Disease - GOLD. 2022 [cited August 21, 2023]. Available in: https://goldcopd.org/2023-gold-report-2/

  8. Global Initiative for Asthma. Global strategy for asthma management and prevention, 2023. Updated May. 2023. Available in: https://ginasthma.org/2023-gina-main-report/

  9. Larenas-Linneman D, Salas-Hernández J, Río-Navarro D, Luna-Pech BE, Navarrete-Rodríguez JA, Gochicoa EM. Manejo integral del asma. Lineamientos para México. Rev Alerg Mex. 2021;68(1):1-122.

  10. Kreitzer SM, Saunders NA, Tyler HR, Ingram RH Jr. Respiratory muscle function in amyotrophic lateral sclerosis. Am Rev Respir Dis [Internet]. 1978;117(3):437-447. Available in: https://pubmed.ncbi.nlm.nih.gov/629478/

  11. Burakgazi AZ, Hoke A. Respiratory muscle weakness in peripheral neuropathies. J Peripher Nerv Syst [Internet]. 2010;15(4):307-313. Available in: http://dx.doi.org/10.1111/j.1529-8027.2010.00293.x

  12. Allen SM, Hunt B, Green M. Fall in vital capacity with posture. Br J Dis Chest [Internet]. 1985;79(3):267-271. Available in: http://dx.doi.org/10.1016/0007-0971(85)90047-6

  13. Qaseem A, Wilt TJ, Weinberger SE, Hanania NA, Criner G, van der Molen T, et al. Diagnosis and management of stable chronic obstructive pulmonary disease: a clinical practice guideline update from the American College of Physicians, American College of Chest Physicians, American Thoracic Society, and European Respiratory Society. Ann Intern Med [Internet]. 2011;155(3):179-191. Available in: http://dx.doi.org/10.7326/0003-4819-155-3-201108020-00008

  14. US Preventive Services Task Force, Mangione CM, Barry MJ, Nicholson WK, Cabana M, Caughey AB, et al. Screening for chronic obstructive pulmonary disease: US Preventive Services Task Force reaffirmation recommendation statement: US preventive services task force reaffirmation recommendation statement. JAMA [Internet]. 2022;327(18):1806-1811. Available in: http://dx.doi.org/10.1001/jama.2022.5692

  15. García-Río F, Calle M, Burgos F, Casan P, Del Campo F, Galdiz JB, et al. Spirometry. Spanish Society of Pulmonology and Thoracic Surgery (SEPAR). Arch Bronconeumol [Internet]. 2013;49(9):388-401. Available in: https://doi.org/10.1016/j.arbres.2013.04.001

  16. Sin DD, Wu L, Man SFP. The relationship between reduced lung function and cardiovascular mortality: a population-based study and a systematic review of the literature. Chest [Internet]. 2005;127(6):1952-1959. Available in: http://dx.doi.org/10.1378/chest.127.6.1952

  17. Qaseem A, Snow V, Fitterman N, Hornbake ER, Lawrence VA, Smetana GW, et al. Risk assessment for and strategies to reduce perioperative pulmonary complications for patients undergoing noncardiothoracic surgery: a guideline from the American College of Physicians. Ann Intern Med [Internet]. 2006;144(8):575-580. Available in: http://dx.doi.org/10.7326/0003-4819-144-8-200604180-00008

  18. Redlich CA, Tarlo SM, Hankinson JL, Townsend MC, Eschenbacher WL, Von Essen SG, et al. Official American Thoracic Society technical standards: spirometry in the occupational setting. Am J Respir Crit Care Med [Internet]. 2014;189(8):983-993. Available in: http://dx.doi.org/10.1164/rccm.201402-0337ST

  19. Davis PB. Pathophysiology of the lung disease in cystic fibrosis. In: Davis PB, editors. Cystic fibrosis. New York: Marcel Dekker; 1993.

  20. Kapnadak SG, Dimango E, Hadjiliadis D, Hempstead SE, Tallarico E, Pilewski JM, et al. Cystic Fibrosis Foundation consensus guidelines for the care of individuals with advanced cystic fibrosis lung disease. J Cyst Fibros [Internet]. 2020;19(3):344-354. Available in: http://dx.doi.org/10.1016/j.jcf.2020.02.015

  21. Hambly N, Farooqi MM, Dvorkin-Gheva A, Donohoe K, Garlick K, Scallan C, et al. Prevalence and characteristics of progressive fibrosing interstitial lung disease in a prospective registry. Eur Respir J [Internet]. 2022;60(4):2102571. Available in: http://dx.doi.org/10.1183/13993003.02571-2021

  22. Finder JD, Birnkrant D, Carl J, Farber HJ, Gozal D, Iannaccone ST, et al. Respiratory care of the patient with Duchenne muscular dystrophy: ATS consensus statement: ATS consensus statement. Am J Respir Crit Care Med [Internet]. 2004;170(4):456-465. Available in: http://dx.doi.org/10.1164/rccm.200307-885ST

  23. Bushby K, Finkel R, Birnkrant DJ, Case LE, Clemens PR, Cripe L, et al. Diagnosis and management of Duchenne muscular dystrophy, part 2: implementation of multidisciplinary care. Lancet Neurol [Internet]. 2010;9(2):177-189. Available in: http://dx.doi.org/10.1016/S1474-4422(09)70272-8

  24. Birnkrant DJ, Bushby KMD, Amin RS, Bach JR, Benditt JO, Eagle M, et al. The respiratory management of patients with Duchenne muscular dystrophy: a DMD care considerations working group specialty article: Duchenne muscular dystrophy. Pediatr Pulmonol [Internet]. 2010;45(8):739-748. Available in: http://dx.doi.org/10.1002/ppul.21254

  25. Bossi G, Cerveri I, Volpini E, Corsico A, Baio A, Corbella F, et al. Long-term pulmonary sequelae after treatment of childhood Hodgkin's disease. Ann Oncol [Internet]. 1997;8(suppl 1):S19-S24. Available in: http://dx.doi.org/10.1093/annonc/8.suppl_1.s19

  26. Cooper BG. An update on contraindications for lung function testing. Thorax [Internet]. 2011;66(8):714-723. Available in: http://dx.doi.org/10.1136/thx.2010.139881

  27. Anaesthetic and respiratory equipment: spirometers intended for the measurement of time forced expired volumes in humans. International Organization for Standardization. 2016. Available in: https://www.iso.org/standard/43761.html

  28. Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, et al. Standardisation of spirometry. Eur Respir J [Internet]. 2005;26(2):319-338. Available in: http://dx.doi.org/10.1183/09031936.05.00034805

  29. Benítez-Pérez RE, Vázquez-García JC, Sánchez-Gallén E, Salas-Hernández J, Pérez-Padilla R, Reyes-Herrera A, et al. Impacto de un programa educativo de espirometría en el primer nivel de atención en México. Neumol Cir Torax [Internet]. 2021;80(1):29-38. Disponible en: https://dx.doi.org/10.35366/99451

  30. Vázquez-García JC, Ortiz-Siordia R, Franco-Marina F, Salas-Hernández J, Benítez-Pérez RE, Pérez-Padilla R. Impacto de un curso taller de cinco horas en la interpretación de la espirometría. Neumol Cir Torax [Internet]. 2019;78(3):270-276. Disponible en: http://dx.doi.org/10.35366/nt193b

  31. Johnson B, Steenbruggen I, Graham BL, Coleman C. Improving spirometry testing by understanding patient preferences. ERJ Open Res [Internet]. 2021;7(1):00712-2020. Available in: http://dx.doi.org/10.1183/23120541.00712-2020

  32. Supplemental Material for: Standardization of Spirometry - 2019 Update: An Official. American Thoracic Society and European Respiratory Society Technical Statement. Available in: https://www.atsjournals.org/doi/suppl/10.1164/rccm.201908-1590ST/suppl_file/graham_data_supplement.pdf

  33. CDC LC. Quick learn: recognize the four biosafety levels [Internet]. Cdc.gov. [cited August 21, 2023]. Available in: https://www.cdc.gov/training/quicklearns/biosafety/

  34. Rasam SA, Apte KK, Salvi SS. Infection control in the pulmonary function test laboratory. Lung India [Internet]. 2015;32(4):359-366. Available in: http://dx.doi.org/10.4103/0970-2113.159571

  35. Franczuk M, Przybylowski T, Czajkowska-Malinowska M, Radlinski J, Bochenek G, Wesolowski S, et al. Spirometry during the SARS-CoV-2 pandemic. Guidelines and practical advice from the expert panel of Respiratory Physiopathology Assembly of Polish Respiratory Society. Adv Respir Med [Internet]. 2020;88(6):640-650. Available in: http://dx.doi.org/10.5603/ARM.a2020.0186

  36. Respiratory function testing during endemic COVID-19 [Internet]. Org.uk. [cited August 22, 2023]. Available in: https://www.artp.org.uk/write/MediaUploads/Standards/COVID19/Respiratory_Function_Testing_During_Endemic_COVID_V1.5.pdf

  37. Milanese M, Corsico AG, Bellofiore S, Carrozzi L, Di Marco F, Iovene B, et al. Suggestions for lung function testing in the context of COVID-19. Respir Med [Internet]. 2020;177:106292. Available in: http://dx.doi.org/10.1016/j.rmed.2020.106292

  38. Saunders MJ, Haynes JM, McCormack MC, Stanojevic S, Kaminsky DA. How local SARS-CoV-2 prevalence shapes pulmonary function testing laboratory protocols and practices during the COVID-19 pandemic. Chest [Internet]. 2021;160(4):1241-1244. Available in: http://dx.doi.org/10.1016/j.chest.2021.05.011

  39. Gochicoa-Rangel L, Torre-Bouscoulet L, Salles Rojas A, Guzmán-Valderrábano C, Silva-Cerón M, Benítez-Pérez RE, et al. Functional respiratory evaluation in the COVID-19 era: the role of pulmonary function test laboratories. Rev Invest Clin [Intertnet]. 2020;73(4). Available in: http://dx.doi.org/10.24875/ric.20000250

  40. Sheikh S, Hamilton FW, Nava GW, Gregson FKA, Arnold DT, Riley C, et al. Are aerosols generated during lung function testing in patients and healthy volunteers? Results from the AERATOR study. Thorax [Internet]. 2022;77(3):292-294. Available in: http://dx.doi.org/10.1136/thoraxjnl-2021-217671

  41. Verbeek JH, Rajamaki B, Ijaz S, Sauni R, Toomey E, Blackwood B, et al. Personal protective equipment for preventing highly infectious diseases due to exposure to contaminated body fluids in healthcare staff. Cochrane Database Syst Rev [Internet]. 2020;5(5):CD011621. Available in: http://dx.doi.org/10.1002/14651858.cd011621.pub5

  42. Crenesse D, Berlioz M, Bourrier T, Albertini M. Spirometry in children aged 3 to 5 years: reliability of forced expiratory maneuvers. Pediatr Pulmonol [Internet]. 2001;32(1):56-61. Available in: http://dx.doi.org/10.1002/ppul.1089

  43. Piccioni P, Borraccino A, Forneris MP, Migliore E, Carena C, Bignamini E, et al. Reference values of forced expiratory volumes and pulmonary flows in 3-6 year children: a cross-sectional study. Respir Res [Internet]. 2007;8(1):14. Available in: http://dx.doi.org/10.1186/1465-9921-8-14

  44. Culver BH, Graham BL, Coates AL, Wanger J, Berry CE, Clarke PK, et al. Recommendations for a standardized pulmonary function report. An official American Thoracic Society technical statement. Am J Respir Crit Care Med [Internet]. 2017;196(11):1463-1472. Available in: http://dx.doi.org/10.1164/rccm.201710-1981ST



AFFILIATIONS

1Instituto Nacional de Enfermedades Respiratorias Ismael Cosío Villegas. Mexico City, Mexico

2Hospital Regional de Alta Especialidad de la Península de Yucatán. Mexico

3Hospital Regional "Dr. Valentín Gómez Farías", ISSSTE. Guadalajara, Jalisco, Mexico

4Atención Integral Respiratoria Especializada. Guadalajara, Jalisco, Mexico

5Instituto de Desarrollo e Innovación en Fisiología Respiratoria. Mexico City, Mexico.



Conflict of interests: the authors declare that they have no conflict of interests.



CORRESPONDENCE

Dra. Laura Gochicoa-Rangel. E-mail: gochis@dr.com




Received: VIII-21-2023; accepted: IX-15-2023.

Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8
Figure 9
Figure 10
Figure 11
Figure 12
Figure 13
Table 1
Table 2
Table 3
Table 4
Table 5
Table 6
Table 7
Table 8
Table 9

2020     |     www.medigraphic.com

Mi perfil

C?MO CITAR (Vancouver)

Neumol Cir Torax. 2023;82