Electronic device to improve the respiratory sinus arrhythmia reproducibility through a close control of the respiratory rhythm
Carrasco SS, Gaitán GMJ, González CR, Yánez SO, Ríos AE
Language: Spanish
References: 15
Page: 61-66
PDF size: 166.93 Kb.
ABSTRACT
Usually, the respiratory frequency during heart rate variability studies is roughly controlled via the metronome sound to keep the respiratory sinus arrhythmia as fixed as possible. Besides a shortage of information, contradictory results about the reproducibility of the respiratory sinus arrhythmia have been reported. Thus, we propose a new electronic device to improve the respiratory sinus arrhythmia reproducibility through a close control of the respiratory rhythm. The device, based on a microcontroller and a column of 60 LED’s display, shows the desired rhythm to be followed by the subject. The instrument was tested three alternate days in 11 healthy subjects breathing at 0.2 Hz during heart rate variability studies, including breathing time series. Temporal and spectral indexes of both time series were derived and their reproducibility evaluated. Intraclass correlation coefficients were from 0.81 to 0.94, which were higher than those reported on the literature. The breathing and heart rate spectra on the high frequency component presented central frequencies of 0.199 ± 0.006 vs 0.199 ± 0.008 Hz (P ‹ 0.05), and bandwidths of 0.013 ± 0.006 vs 0.016 ± 0.008 Hz (P › 0.05), respectively. Since the device induced similar central frequencies, small bandwidths, and the reproducibility was higher than previously reported, we assume that the influence of the device was acceptable.
REFERENCES
Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. “Heart rate variability: standards of measurement, physiological interpretation, and clinical use”. Circulation 1996; 93: 1043-1065.
Bernstson GG, Bigger JT, Eckberg DL, Grossman P, Kaufmann PG, Malik M et al. “Heart rate variability: origins, methods, and interpretative caveats”. Psychophysiology 1997; 34: 623-648.
Malik M. “Heart rate variability”. Curr Opinion Cardiol 1998; 13: 36-44.
Angelone A, Coulter N. “Respiratory sinus arrhythmia: A frequency dependent phenomenon”. J Appl Physiol 1964; 19: 479-482.
Saul JP, Cohen RJ. “Respiratory sinus arrhythmia”. In: Levy MN, Schwartz PJ, Editores, Vagal Control of the Heart: Experimental Basis and Clinical Implications, Futura Publishing Co. (EU), 1994: 511-535.
Frederiks J, Swenne C, TenVoorde B, Honzikova N, Levert J, Maan A et al. “Non-baroreflex mediated heart rate variability causes overestimation of baroreflex sensitivity”. Comput Cardiol 1997; 24: 199-202.
Grossman P, Karemaker J, Wieling W. “Prediction of tonic parasympathetic control using respiratory sinus arrhythmia: The need for respiratory control”. Psychophysiology 1991; 28: 201-216.
Kleiger R, Stein P, Bosner M, Rottman J. “Time domain measurements of heart rate variability”. Cardiol Clin 1991; 10: 487-498.
Raymond B, Mazumdar J, Nadagopal D. “Modeling the shift of the low frequency component of the heart rate variability”. Comput Cardiol 1997; 24: 407-410.
Cohen L. “The time and frequency description of signals” In: Time-frequency analysis, Prentice-Hall (EU), 1995: 1-26.
Pitzalis MV, Mastropasqua F, Massari F, Forleo C, Di Maggio M, Passantino A et al. “Short- and long-term reproducibility of time and frequency domain heart rate variability measurements in normal subjects”. Cardiovasc Res 1996; 32: 226-233.
Strauss-Blasche G, Moser M, Voica M, Mcleod D, Klammer N, Marktl W. “Relative timing of inspiration and expiration affects respiratory sinus arrhythmia”. Clin Exp Pharmacol Physiol 2000; 27: 601-606.
Stein P, Rich M, Rottman J, Kleiger R. “Stability of index of heart rate variability in patients with congestive heart failure”. Am Heart J 1995; 129: 975-981.
Sinnreich R, Kark J, Friedlander Y, Sapoznikov D, Luria M. “Five-minute recordings of heart rate variability for population studies: repeatability and age-sex characteristics”. Heart 1998; 80: 156-162.
Lawrence GP, Home PD, Murray a. “Repeatability of measurements and sources of variability in test of cardiovascular autonomic function”. Br Heart J 1992; 68: 205-211.