medigraphic.com
SPANISH

Revista Mexicana de Ingeniería Biomédica

  • Contents
  • View Archive
  • Information
    • General Information        
    • Directory
  • Publish
    • Instructions for authors        
  • medigraphic.com
    • Home
    • Journals index            
    • Register / Login
  • Mi perfil

2000, Number 4

<< Back

Rev Mex Ing Biomed 2000; 21 (4)

Monitoring and control system for temperature that will use like reference for the calibration of the viscosimeters at National Center of Metrology

Blancas J, Conejo A
Full text How to cite this article

Language: Spanish
References: 7
Page: 142-147
PDF size: 100.80 Kb.


Key words:

, Monitoring and control system for temperatura, Capilaries viscosimeters and the wide of pulse technique.

ABSTRACT

We have developed a monitoring and control system for temperature of water bath in which required a temperature of 293.150 K +/- 5 mK, this uncertainty is necesary so that it can work like national patron. Inside of this bath will introduced the masters capilaries viscosimeters that will use like reference for the calibration of the viscosimeters that will arrived to National Center of Metrology. The control of temperature is made by electrical resistor of 1500 W. This resistor’s heat is made electrically by using the wide of pulse technique (PWM).


REFERENCES

  1. B. Charles. Enciclopedia de la Electrónica, Editorial Océano, 1995. Tomos 4, 5 y 6.

  2. Chester LN. Instrumentation and control, Wiley Interscience, 1990.

  3. David W. ST. Clair. Controller tuning and control loop performance, second edition, Straight-line control company incorporated,1995.

  4. González VJA. Introducción a los microcontroladores, McGraw-Hill, 1992.

  5. Hugh, Hildreth, Skilling, Circuitos en ingeniería eléctrica, Edit. C.E.C.S.A., 1990.

  6. Gualda JA, Martínez S, Martínez PM, Electrónica Industrial: Técnicas de Potencia. 2a Edición, Editorial Alfaomega Marcombo, 1995.

  7. National Instruments, LabView user manual, 1996.




2020     |     www.medigraphic.com

Mi perfil

C?MO CITAR (Vancouver)

Rev Mex Ing Biomed. 2000;21