Rev Cubana Invest Bioméd 2012; 31 (1)

Reference values for electrooculography

Barrientos CA, Herrera MM, García BO, Sierra HM, Pérez DO
Full text How to cite this article

Language: Spanish
References: 11
Page: 46-52
PDF size: 121.48 Kb.


Key words:

electrooculogram, retinal pigment epithelium, reference values.

ABSTRACT

Objective: obtain electrooculographic reference values based on the patterns set by the Standardization Committee of the International Society for Clinical Electrophysiology of Vision (ISCEV).
Methods: twenty-seven healthy individuals were selected at random among employees at "Camilo Cienfuegos" International Center for Retinitis Pigmentosa. An electrooculogram was performed on them following the parameters set by ISCEV Standardization Committee. A measurement was made of the main clinically important electrooculographic indicators (Arden index, dark peak amplitude and light peak emergence time in the light phase).
Results: the lowest amplitude values of the potential ranged between 388 and 882 µv in the dark phase. The light peak was obtained between 9 and 10 minutes, and during this phase the potential reached an amplitude ranging between 808 and 1 963 µv. This amplitude variability may be related to the fact that the test was conducted without pupillary mydriasis. The reference value obtained for Arden index was 1,55 to 2,87.
Conclusions: following standardized methodologies, reference values were obtained for the main clinically useful electrooculographic indicators. These indicators will be of use in our laboratory.


REFERENCES

  1. Fishman GA. The electro-oculogram. En: Fishman GA. Bich DG, Holder GE. Electrophysiologic Testing in disorders of retina, optic nerve and visual pathway. Ophthalmology monographs 2 of American Academy of Ophthalmology. 2nd ed. San Francisco: The Foundation of the American Academy of Ophthalmology; 2001. p. 157-75.

  2. Kolder HE. Electro-oculography. En: Heckenlively JR, Arden GB. Principles and practice of clinical electrophysiology of vision. St. Louis: Mosby Year Book; 1991. p. 301-13.

  3. Miyake Y. Electrooculography. En: Miyake Y. Electrodiagnosis of retinal diseases. New York: Springer-Verlag; 2006. p. 165.

  4. Marmorstein LY, Wu J, McLaughlin P, Yocom J, Kart MO, Neussert R, et al. The light peak of the electroretinogram is dependent on voltage-gated calcium channels and antagonized by bestrophin (Best-1). Gen Physiol. 2006 May;127(5):577-89.

  5. Farrell DF. Unilateral retinitis pigmentosa and cone-rod dystrophy. Clin Ophthalmol. 2009;3:26370.

  6. Sohn EH, Francis PJ, Duncan JL, Weleber RG, Saperstein DA, Farell DF, et al. Phenotypic variability due to a novel Glu292Lys variation in exon 8 of the BEST1 gene causing best macular dystrophy. Arch Ophthalmol. 2009 July;127(7):913-20.

  7. Spaide RF, Noble K, Morgan A, Freund KB. Vitelliform macular dystrophy. Ophthalmology. 2006 Aug;113(8):1392-400.

  8. Marmor MF, Zrenner E. Standard for clinical electro-oculography. Doc Ophthalmol. 1993;85:115-24.

  9. Marmor MF. Standardization notice: EOG standard reapproved. Doc Ophthalmol. 1998;95:912.

  10. Brown M, Marmor M, Vaegan, Zrenner E, Brigell M, Bach M. ISCEV Standard for clinical electro-oculography (EOG). Doc Ophthalmol. 2006;113:205-12.

  11. Benítez del Castillo JM, Pérez JL, Benítez del Castillo J, Pérez E. Potencial de reposo de la retina. Electro-oculograma. En: Benítez del Castillo JM, Pérez JL, Benítez del Castillo J, Pérez E. Manual Básico de electrofisiología ocular sus aplicaciones en la clínica. Jerez de la Frontera: MAC LINE, S.L; 2002. p. 27-36.




2020     |     www.medigraphic.com

Mi perfil

C?MO CITAR (Vancouver)

Rev Cubana Invest Bioméd. 2012;31