medigraphic.com
SPANISH

Revista Cubana de Farmacia

ISSN 1561-2988 (Print)
  • Contents
  • View Archive
  • Information
    • General Information        
    • Directory
  • Publish
    • Instructions for authors        
  • medigraphic.com
    • Home
    • Journals index            
    • Register / Login
  • Mi perfil

2015, Number 4

<< Back Next >>

Rev Cubana Farm 2015; 49 (4)

Validation of the enzymatic method for the estimation of serum and urine creatinine

Carrión DIR, García BL, Suárez PY, Rodríguez FB, Aja MG
Full text How to cite this article

Language: Spanish
References: 13
Page: 618-629
PDF size: 189.21 Kb.


Key words:

creatinine, spectrophotometry, clinical laboratory, validation.

ABSTRACT

Introduction: creatinine is an important control and monitoring marker in various kidney pathologies.
Objectives: to validate a colorimetric enzymatic method of kinetic analysis developed with Cuban-made reagents, in order to determine the serum and urine creatinine for future application in the clinical laboratories.
Methods: the enzymatic method is used for kinetic determinations based on the spectrophotometric measurement at 550nm wavelength of the formed quinoneimine. The validation evaluated the parameters specificity, linearity, precision, accuracy, detection and quantitation limits in compliance with the present Cuban regulations.
Results: the method was specific for the suggested objective; provided linear response from 26,52 to 2 652 µmol/L in serum and from 884 to 26 520 µmol/L in urine; had great precision and accuracy and its detection and quantitation limits were adequate.
Conclusions: since the evaluated method fulfilled the present regulatory demands, it may be used in the domestic clinic laboratories to determine serum and urine creatinine.


REFERENCES

  1. Burtis CA, Ashwood ER. Tiezt Texbook of Clinical Chemistry. Third Edition. Philadelphia: WB Saunders Co ed; 2001. p. 1241-45.

  2. Gurdián M, Kontorovsky I, Alvarado E, González A. Manual de procedimientos de labotarorio en bioquímica clínica y control de calidad. Managua: Ministerio de Salud de Nicatagua; 2004. p85-88.

  3. Aja G. Desarrollo de un diagnosticador para determinar Creatinina en suero. Tesis de Maestría de Tecnología y Control de Medicamentos. La Habana: IFAL. UH; 2007.

  4. Perazzi B, Angerosa M. Creatinina en sangre: calidad analítica e influencia en la estimación del Índice de Filtrado Glomerular. Acta Bioquím. Clín. Latinoam. 2011;45(2):5-7.

  5. Myers G. Recommendations for improving serum measurement: A report from the laboratory working group of the National Kidney Disease Education Program. Clin. Chem. 2006;52(5):101-5.

  6. Roberts WL. Reference information for the Clinical Laboratory/ In: Burtis CA, Ashwood ER, Bruns DE (Eds). Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. Fourth Edition. St. Louis, Missouri: Elsevier Inc; 2006. pp. 2251-2318.

  7. Murray RL. Creatinine. In: Pesce-Kaplan Publisher. Laboratory Medicine: A Scientific and Managerial Infobase. [CD-ROM] V. 4.2. Cincinati: Pesce-Kaplan Publisher; 2000.

  8. American Association for Clinical Chemistry (AACC). Kidney and urinary tract function, disorders and diseases. March 7, 2002. [citado: 15 Sep 2014] Disponible en: http://labtestsonline.org/understanding/conditions/

  9. Greenberg N, Roberts WL, Bachmann LM, Wright EC, Neil DR, Jack J, et al. Specificity Characteristics of 7 Commercial Creatinine Measurement Procedures by Enzymatic and Jaffe Method Principles. Clinical Chemistry. 2012;58(2):391-401.

  10. Marakala V, Avinash SS, Shivashankara AR, Malathi M, Arun K. Serum creatinine assay: enzymatic vs kinetic jaffe's method. Journal of Evolution of Medical and Dental Sciences. 2012 Oct-Dec;1(4):258-264.

  11. Çuhadar S. Preanalytical variables and factors that interfere with the biochemical parameters: a review. OA Biotechnology. 2013 Jun 01;2(2):19.

  12. Sugita O, Uchiyama K, Yamada T, Sato T, Okada M, Takeuchi K. Reference values of serum and urine creatinine, and of creatinine clearance by a new enzymatic method. Ann Clin Biochem. 1992;29 (Pt. 5):523-528.

  13. Barreto Penie J, Santana Porben S, Consuegra SD. Intervalos de referencia locales para la excreción urinaria de creatinina en una población adulta. Nutr. Hosp. 2003;XVIII(2):65-75.




2020     |     www.medigraphic.com

Mi perfil

C?MO CITAR (Vancouver)

Rev Cubana Farm. 2015;49