Table 6: Acceptability criteria for

measurement of intrathoracic gas volume

(functional residual capacity).

Tidal volume breathing before shutter closure and gasping

Acceptable

Before shutter closure:

• Stable end-expiratory tidal volume

During shutter closure:

• Gasping maneuvers

• Superimposed straight lines without hysteresis

• Parallel straight lines with minimal hysteresis

• Gasping respiratory rate between 0.5 to 1 Hz or between 1.0 and 1.5 Hz with normal spirometry or mild obstruction

Usable

Any of:

Before shutter closure:

• End-expiratory tidal volume unstable, but no significant change in either direction

During shutter closure:

• Partially closed gasping maneuver

• Partially overlapping straight lines

• Parallel straight lines with hysteresis

• Respiratory rate >1.5-2.0 Hz with normal spirometry or mild obstruction

Not

acceptable

or rejected

Any of:

Before shutter closure:

• Unstable end-expiratory tidal volume with significant change in either direction (e.g., increase in end-expiratory lung volume with each breath)

During shutter closure:

• Gasping maneuver with mouth opening

• No straight lines

• Excessive hysteresis

• Force to perform the gasping maneuver exceeded the range of the oral pressure transducer

• Gasping respiratory rate < 0.5 Hz, > 2 Hz with normal spirometry or mild obstruction or > 1.5 with significant obstruction