Cardiac pacemaker dysfunction due to rotation of the battery (reel syndrome)

Guevara-Valdivia ME, Leal-Ortiz GE, Bonilla-Morales I
Full text How to cite this article

Language: Spanish
References: 8
Page: 209-212
PDF size: 62.68 Kb.


Key words:

pacemaker artificial, equipment failure, electrodes implanted, foreign-body migration, reel syndrome.

ABSTRACT

Reel syndrome is characterized by the rotation of permanent pacemaker generator on its transverse axis and electrode catheters curl around it, so this causes displacement of the electrodes with the loss of atrial and ventricular pacing. It can cause symptoms due to dysfunction of the pacemaker and can be serious even those patients dependent on the pacemaker. We present three patient cases who after the placement of pacemaker attended by disfunction of the displacement of the electrodes their clinical pictures were compatible with reel syndrome.


REFERENCES

  1. Femenía FJ, Arce M, Peñafort F, Arrieta M, Gutiérrez D. Complicaciones del implante de marcapaso definitivo. ¿Un evento operador dependiente? Análisis de 743 pacientes consecutivos. Arch Cardiol Mex 2010;80(2):95-99. Disponible en http://www.scielo.org.mx/scielo.php?script=sci_ arttext&pid=S1405-99402010000200006

  2. Carnero-Varo A, Pérez-Paredes M, Ruiz-Ros J. A, Giménez- Cervantes D, Martínez-Corbalán FR, Cubero-López T, et al. Reel syndrome: a new form of Twiddler’s syndrome. Circulation 1999;100(8):e45-e46. Disponible en http://circ. ahajournals.org/content/100/8/e45.full.pdf+html

  3. Vural A, Agacdiken A, Ural D, Komsuoglu B. Reel syndrome and pulsatile liver in a patient with a two-chamber pacemaker. Jpn Heart J 2004;45(6):1037-1042. Disponible en http://www. jstage.jst.go.jp/article/jhj/45/6/1037/_pdf

  4. Aliyev F, Celiker C, TürkoGlu C, Turhan FN. Early development of pacemaker Reel syndrome in an elderly patient with cognitive impairment. Turk Kardiyol Dern Ars 2009; 37(7):488-489.

  5. Patel MB, Pandya K, Shah AJ, Lojewski E, Castellani MD, Thakur R. Reel syndrome-not a twiddler variant. J Interv Card Electrophysiol 2008;23(3):243-246.

  6. Ejima K, Shoda M, Manaka T, Hagiwara N. Reel syndrome. J Cardiovasc Electrophysiol 2009;20(7):822.

  7. Gámez-López AL, Bonilla-Palomas JL, Granados AL. An unusual case of cardiac resynchronization therapy nonresponder: the reel syndrome. Europace 2010;12(6): 778. Disponible en http://europace.oxfordjournals.org/content/ 12/6/778.full.pdf+html

  8. Vraziæ H, Schulz S, Kolb C. ´Selective´ reel syndrome. Europace 2011;13(1):143. Disponible en http://europace. oxfordjournals.org/content/13/1/143.1.long